103N / 105N / 106
EN
Overload Protection : 1A , 600V IR 10KA fuse (Bussmann
BBS-1 or equivaknt) for mA input. (Size 10.3mm x 34.9mm)
15A , 600V IR 100KA fuse (Bussman KTK 15 or equivalent)
for A input. (Size 10.3mm x 38.1mm)
AC Conversion Type : AC conversions are average
sensing, calibrated to the rms value sine wave input.
(5) Resistance
Range Resolution
Accuracy
103N 105N / 106
±(0.7%reading +
3digits)
±(0.7%reading +
3digits)
±(1.5%reading +
5digits)
±(0.6%reading +
3digits)
±(1.5%reading +
5digits)
400Ω 0.1Ω
40KΩ 10Ω
4KΩ
400KΩ 100Ω
4MΩ
40MΩ
1KΩ
10KΩ
±(0.7%reading +
3digits)
±(0.6%reading
+3digits)
±(0.4%reading +
2digits)
Over voltage
protection
600V rms
Open circuit Voltage : -1.3V approx.
(6) Diode Check and Continuity
Range
Resolution Accuracy
Max. Test Current
Current
Max. Open Circuit
Circuit Voltage
±(1.5%reading +
5digits)*
1mV 1.5mA 3V
* For 0.4V — 0.8V.
Overload Protection : 600V rms max.
Continuity : The internal sounder operates when resistance
is less than 30Ω approximately.
(7) Frequency / RPM
Range Resolution
Accuracy
Sensitivity
4.0KHz/40KRPM
40KHz/400KRPM
400KHz/4MRPM
4MHz/40MRPM 1KHz/30KRPM 300mV rms
40MHz/
400MRPM
10KHz/
300KRPM
1V rms
* 400MHz/
4000MRPM
100KHz/
3MRPM
* *
1Hz/30RPM
10Hz/300RPM
100Hz/3KRPM
150mV rms ≥ 20Hz
1.5V rms ≤ 20Hz
Frequency :
±(0.01%reading
+ 1digit)
RPM :
±(0.01%reading
+ 10digits)
600V rms
Over voltage
protection
8