Electromagnetic Compatibility16
Electromagnetic Compliance
The KWIKtrak Gate has been tested for compliance with current regulatory standards regarding
its capacity to block EMI (electromagnetic interference) from external sources.
Nonetheless, some procedures can help reduce electromagnetic interferences:
(QVXUHWKDWRWKHUGHYLFHVLQSDWLHQWPRQLWRULQJDQG௘௘RUOLIHVXSSRUWDUHDVFRPSO\WR
accepted emissions standards.
Maximize the distance between electro-medical devices. High-powered devices may
SURGXFH(0,WKDWFDQDႇHFWWKHSURGXFW
For more information on how to manage the unit’s RF electromagnetic environment, please
consult the AMI TIR 18-1997 - Guidance on Electromagnetic Compatibility of Medical Devices for
&OLQLFDO௘௘%LRPHGLFDO(QJLQHHUV
WARNING8VHRIDFFHVVRULHVFDEOHVDQGVSDUHSDUWVRWKHUWKDQWKRVHVSHFL¿HGRUSURYLGHGE\$UMR
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
WARNING8VH RIWKLVHTXLSPHQW DGMDFHQW WRRUVWDFNHG ZLWKRWKHUHTXLSPHQW VKRXOGEHDYRLGHG
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING: The equipment may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take action, such as reorienting, relocating the equipment or shielding
the location.
Electromagnetic Emissions
Guidance and Manufacturers Declaration -
Electromagnetic Emissions - For all Equipment and Systems
The KWIKtrak Gate is intended for use in the electromagnetic environments indicated below.
The customer or the user of the KWIKtrak Gate should make sure that it is used in such
environments.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The KWIKtrak Gate uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The KWIKtrak Gate is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
ÀXFWXDWLRQV௘௘ÀLFNHU
emissions
IEC 61000-3-3
Not applicable
Electromagnetic Compatibility