SYMPTOM POSSIBLE CAUSE REMEDY
After using for a while,
push buttons can operate
normally but no air is
coming out.
Blower over heated.
(If you switch off the Air Supply
Unit for 30 minutes and turn
the Unit back on, unit is able to
operate normally again).
Check Heat Exhaust
hose.
Clean or replace air
lter.
Contact Arjo Service
Representative.
Blower still runs but cannot
change between modes.
(i.e. cannot change LED
lights from INSTAFLATE to
STATIC).
CPU needs to be reset. (If you
turn master switch off for 10
seconds, and back on, unit
should revert back to normal).
If this happens on the
same machine / unit
twice, contact Arjo
Service Representative.
Alarm Sounds. Power to unit has been
interrupted or turned off.
Push Alarm Reset
button on front of unit,
and / or re-apply power.