Always record your results in your self-testing logbook along with
other information such as insulin dosage, diet, and exercise.
The result is automatically stored in memory with the time and date.
Pick up the meter. Touch the end of the test strip
to the blood sample until the reaction site is full.
Immediately remove the meter and test strip from
the blood drop.
The meter will count down and display result in 7 seconds.
CONSIDER YOUR RESULT.
REPEAT ALTERNATE SITE TEST:
• IF the blood sample appeared to be diluted with clear uid.
• IF you did not vigorously rub the test site.
• IF the blood drop was not large enough to ll the reaction site
• IF the test was accidentally marked as a “Control” result.
• IF your result was not consistent with how you feel.
• IF more than 20 seconds elapsed from sample collection to
measurement (evaporation of the blood sample may cause
a test result that is higher than the actual value).
Any of the above situations can lead to an inaccurate test result.
If the repeated alternate site result is still not consistent with how you
feel, conrm your blood glucose level with ngertip testing.
IMPORTANT!
Insert strip
VERTICALLY into
blood sample,
NOT at an angle.
Remove test strip from the
bottle. Immediately replace
the bottle cap tightly. Insert
test strip as shown in diagram.
A beep will sound and the
meter will turn on automatically.
Verify that all symbols appear
on the screen.
Verify that the code number
displayed on the screen
matches the code number on
the bottle of test strips used.
Performing a Blood Glucose Test From an Alternate Site
Step 5
Step 6
Step 7
Step 8
Remove the test strip and dispose of it in accordance with local
guidelines or as directed by your healthcare professional.
The meter shuts off automatically when the test strip is removed.
Code No.
(F-2, 3 or 4)
Expiration date
F- 4
7
0018
F-4
Choose a Site: Select a soft,
eshy area on the palm that is
free of visible veins and moles
and away from bone. Wash the
area with warm, soapy water.
Rinse and dry completely. If you
use alcohol wipes to cleanse
the site, make sure the area is
dry before lancing the site.
Step 1
Step 3
NOTE
If it took longer than 20 SECONDS to
obtain a blood sample and touch the
strip to the blood drop, repeat test.
Consider AST when: Use fingertip testing:
• Testing before a meal. • Within two hours after a
meal.
• You are in a fasting state. • Within two hours after insulin
dosing.
• Two hours have passed • Within two hours after
since a meal. physical activity.
• Two hours have passed • If you have a history of
since insulin dosing. hypoglycemia, are
experiencing low blood
• Two hours have passed glucose, or suffer from
since physical activity. hypoglycemic unawareness
(you cannot tell when you
have low blood glucose).
• To conrm your blood
glucose level if your AST
result is not consistent
with how you feel.
• During times of stress or
illness.
• Was the sample free of clear or watery uid?
• Was the drop large enough (at least 0.3 µL)?
• Did the reaction site ll with enough blood all at once?
IF NOT,
REPEAT TEST.
IMPORTANT
Filled
Not Filled
d) Once a large enough drop of blood has
formed, remove the lancing device.
Insert lancet and cock
lancing device.
Step 4
b) Firmly hold the cocked lancing device
against the clean skin for 5-10 seconds.
a) PRESS AND VIGOROUSLY RUB THE SELECTED
AREA for 10 SECONDS until it starts to feel
warm to the touch.
Step 2
a
b
c
d
IMPORTANT
Repeat blood draw if uid is clear.
c) Press the release button
on the lancing device
to lance the skin. Continue
to hold the lancing device
firmly against the skin until
a blood drop forms.

Other manuals for Arkray Glucocard 01

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray Glucocard 01 and is the answer not in the manual?

Arkray Glucocard 01 Specifications

General IconGeneral
BrandArkray
ModelGlucocard 01
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals