Step 1: Remove Assure
®
Platinum Test Strip from the bottle.
Immediately replace the bottle cap tightly.
Insert test strip as shown in the diagram. Meter will turn on automatically.
Verify that all symbols appear on the screen.
The strip and flashing blood drop indicate the meter is ready for use.
Step 2: Obtain a blood sample. Touch the sample application tip at the end
of the strip to the blood drop on your finger. Capillary action will pull
the blood into the strip.
Note: Apply blood to test strip within 20 seconds of obtaining blood sample. The
meter will start to count down when it detects that blood has been applied. If meter
does not start after adding the first blood drop, remove the test strip and test with a
new test strip. Discard strip if meter does not start to count down.
Step 3: Meter will count down by showing “7”.
The meter will show the result in 7 seconds.
Option: Flagging Test Result in Memory
Results may be marked with an “!” if you feel the result is incorrect or the test
procedure was not followed correctly.
Mark the test result with an “!” by pressing either the Forward or Back button
after result is on screen. Marked results will not be included in the 7-, 14-, 30-, or
90-day averages.
Healthcare
Professionals: Record test result in patients chart.
mg/dL
am
mg/dL
am
PERFORMING A BLOOD GLUCOSE TEST
Caution: Failure to follow instructions may cause inaccurate results.
Note: If both PCS and
am
appear on the screen, perform a control
solution test (see back).
Results will be shown in units of mg/dL.
Results will not show a decimal point.
Important:
The minimum sample volume is 0.5 µL. Smaller samples may not start the meter or
can start the meter count down and provide an E13 error message. Ensure blood
completely fills the reaction site.
Insert contact bars
into meter.
Sample Application Tip
Contact
Bars
am
mg/dL
Step 4: The flashing arrow reminds you to remove the test strip. Remove the test strip by holding the
meter so that the test strip points down.
Push the test strip release button forward. Dispose of
test strip. The meter shuts off automatically when you remove the test strip.
Enough
Blood
Not Enough
Blood
Too Much
Blood
mg/dL
PROFESSIONAL BLOOD GLUCOSE MONITORING SYSTEM
ASSURE
®
PLATINUM

Other manuals for Arkray ASSURE PLATINUM

Question and Answer IconNeed help?

Do you have a question about the Arkray ASSURE PLATINUM and is the answer not in the manual?

Arkray ASSURE PLATINUM Specifications

General IconGeneral
Measurement Range20-600 mg/dL
Sample Size0.5 μL
Test Time5 seconds
DisplayLCD
Power Source2 x AAA batteries
Operating Temperature10-40°C
Operating Humidity10-90% RH
Measurement MethodElectrochemical
Memory CapacityUp to 500 test results with date and time
Weight50 g

Summary

Performing a Blood Glucose Test

Remove and Insert Test Strip

Remove strip, replace cap, insert into meter, verify symbols.

Obtain Blood Sample

Touch sample tip to blood drop, capillary action pulls blood.

Meter Countdown and Result Display

Meter counts down, shows result in 7 seconds, option to flag.

Remove Test Strip After Use

Remove test strip as indicated, meter shuts off automatically.

Performing a Control Solution Test

Insert Test Strip for Control Test

Insert strip, meter turns on, check symbols.

Enter Control Solution Mode

Press button to enter mode for validating results.

Prepare and Apply Control Solution

Mix solution, discard first drop, apply second drop.

Compare Control Result to Range

Compare result to bottle range to ensure accuracy.

Remove Test Strip After Control

Remove test strip, meter shuts off.

QC Reminders and Troubleshooting

24-Hour QC Reminder

Reminder to perform control solution test.

Low Glucose Level Alert

Alert for < 20 mg/dL, retest or consult doctor.

High Glucose Level Alert

Alert for > 600 mg/dL, retest or consult doctor.

Battery Replacement and Date/Time Setting

Set time/date after battery replacement.

Error Codes E01 and E13

Errors for used strip, meter reading problem, or insufficient blood.