Artel PCS - Entering Calibration Data

Artel PCS
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The Instrument will prompt to enter an operator ID
(Figure 6). Enter a number (not exceeding 20 digits) to
identify the operator. For alpha character entry, refer to
Appendix E of the PCS Procedure Guide.
Enter the pipette ID (Figure 7). Type in a number not
exceeding 20 digits and press ENTER. The printer will
print the calibration header information.
Enter the volume (in microliters) to be dispensed and
press ENTER (Figure 8). The CLEAR key can be used
to correct any typing errors.
Enter Operator ID:
Enter Pipette ID:
Enter Pipette
Volume: ___ µL
Figure 6
Figure 7
Figure 8
PCS Quickstart-15A3179_PCS Quickstart 15A3179 Rev 4/3/13 2:02 PM Page 10