EasyManua.ls Logo

Artel PCS - Inserting Blank Vial

Artel PCS
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Always refer to the PCS Procedure Guide for further instruction
Important Notes:
Use the CAL A vial from the Reagent Kit and not from the Instrument Calibrator Kit.
Never remove the cover of the CAL A vial.
The label for the CAL A vial should face the front of the Instrument with the vertical line facing forward.
Each Blank vial contains a predetermined volume of solution. Any spillage will reduce accuracy.
The lower portion of the Blank vial must be free of smudges. Touching this area should be avoided.
Once the Blank vial has been inserted into the Instrument, it must not be turned or moved until the calibration
procedure is complete for that vial.
The Instrument will then prompt to insert a blue Blank
vial into the vial holder (Figure 5). Remove the Blank vial
from the vial block and remove the cap.
Insert the open vial into the vial holder pressing down
until it contacts the bottom of the vial holder.
Close the Instrument cover. The Instrument will read the
Blank vial.
Remove Cap and
Insert Blank
Figure 5
PCS Quickstart-15A3179_PCS Quickstart 15A3179 Rev 4/3/13 2:02 PM Page 9

Related product manuals