Starting a Pipette Calibration
At the Main Menu prompt (Figure 1), press YES to
begin the pipette calibration procedure. The Instrument
will perform a temperature check.
The Instrument will prompt for lot code (Figure 2). The
Instrument retains up to 6 previously entered lot codes.
If the entered code matches one previously entered,
press NO and re-enter the code.
Press YES if entering a new lot code. Group numbers
found on the correct side of the lot code card must be
entered. Press ENTER after each group of digits is
entered.
A confirmation prompt will be displayed (Figure 3).
Press YES.
The Instrument will perform a brief self-calibration.
Immediately following the self-calibration the Instrument
will prompt to insert the CAL A vial (Figure 4).
Open the vial holder cover and insert the CAL A vial
pressing down until it meets the bottom of the vial
holder.
The Instrument will then perform a zero check and
prompt to remove the CAL A when the test is complete.
Return the CAL A vial to the Reagent Kit for future use.