Always refer to the PCS Procedure Guide for further instruction
Important Notes:
The proper range of Sample Solution for the pipette volume to be tested must be used.
Each Reagent Kit contains bottles of Sample Solution corresponding to various pipette volume ranges.
If an improper Sample Solution has been dispensed into the Blank vial, the results obtained will not be valid and
the vial is no longer usable. It must be discarded.
The number of samples that can be dispensed in this
vial, at the selected volume, will be displayed
(Figure 9).
Press YES and the Instrument will display the
appropriate Sample Solution to use for the pipette
calibration.
The Instrument will take a final reading of the Blank vial
and then prompt for the first dispense of the Sample
Solution with the pipette under test (Figure 10).
XX Left this Vial
Proceed? (Yes/No)
Range X
Pipette Sample #1
Figure 9
Figure 10
Pipette
Calibration
Setup