Always refer to the PCS Procedure Guide for further instruction
Important Notes:
• The proper range of Sample Solution for the pipette volume to be tested must be used.
• Each Reagent Kit contains bottles of Sample Solution corresponding to various pipette volume ranges.
• If an improper Sample Solution has been dispensed into the Blank vial, the results obtained will not be valid and
the vial is no longer usable. It must be discarded.
• The number of samples that can be dispensed in this
vial, at the selected volume, will be displayed
(Figure 9).
• Press YES and the Instrument will display the
appropriate Sample Solution to use for the pipette
calibration.
• The Instrument will take a final reading of the Blank vial
and then prompt for the first dispense of the Sample
Solution with the pipette under test (Figure 10).
XX Left this Vial
Proceed? (Yes/No)
Range X
Pipette Sample #1
Figure 9
Figure 10
Pipette
Calibration
Setup
PCS Quickstart-15A3179_PCS Quickstart 15A3179 Rev 4/3/13 2:02 PM Page 11