EasyManua.ls Logo

Artel PCS - Reading and Result Display

Artel PCS
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Always refer to the PCS Procedure Guide for further instruction
Mixing Sample #1
Please Wait
Reading Sample #1
Please Wait
Result #1 XXX.X
Pipette Sample #2
Rules for Adjustable Volume Pipettes:
Calibration can be performed by starting at any volume.
When the desired number of samples at the first volume have been dispensed, press ENTER.
If a data point must be deleted, press YES at the reprint results prompt and follow the delete a data point instructions.
Enter the next volume in the pipettes range. This process can be repeated. Press END OF RUN when complete.
In some cases, more than one Sample Solution will be required. Special attention should be paid to the screen
prompts to ensure usage of correct solution.
Reading the Sample
Once the sample is dispensed, and the cover is closed, the
Instrument will mix the sample and display a prompt (Figure 11).
The Instrument will then read the absorbance of the sample
and display a prompt (Figure 12).
Finally, the result of the reading will be displayed on the screen
and the Instrument will prompt for the dispense of the next
sample (Figure 13).
Follow the same steps listed on this page for the next sample
and every subsequent sample.
When performing a single volume calibration and the desired
number of data points have been collected, press END OF
RUN. When performing a multi-volume calibration, press
ENTER (see shaded section below).
The group statistics will be computed and the results printed.
Figure 11
Figure 12
Figure 13
Pipette
Calibration
PCS Quickstart-15A3179_PCS Quickstart 15A3179 Rev 4/3/13 2:02 PM Page 13

Related product manuals