Do you have a question about the Astell AMB230 and is the answer not in the manual?
Describes safety compliance with IEC61010 standards.
Explanation of common safety and operational symbols used in the manual.
Details electrical hazards and fire risks associated with the equipment.
Outlines the recommended schedule for periodic maintenance and safety valve testing.
Specifies the intended applications and purpose of the steriliser.
Details the design features and measures implemented to reduce operational risks.
Explanation of internal fuses, circuit protection, and trip mechanisms.
Recommended frequency for performing safety valve tests.
Instructions on how to access and run the safety valve test program.
A concise step-by-step guide for performing the safety valve test.
Defines the intended and approved applications for the steriliser.
How alarm conditions are reported, displayed, and cleared by the system.
Details how the system behaves after power interruptions during or before a cycle.
Indicates a successful cycle completion and related safety checks.
Guidance on assessing load safety and sterility after a cycle.
Options for re-sterilizing the load if the cycle was incomplete or failed.
Indicates a failed cycle and safety precautions for opening the door.
Safety protocols if the door remains locked after a cycle failure.
Initial configuration steps for enabling CFR user logging and audit trails.
Detailed process for adding new user accounts to the system.
Steps to create a backup copy of the audit trail data onto a USB drive.
Procedure for manually advancing or skipping stages within a cycle.
Explanation of the safety interlock that prevents opening when the chamber is pressurized.
Description of the system that shuts down heating upon detecting overpressure.
Details various temperature-based safety interlocks and their functions.
Mechanisms that prevent premature door opening during cooling phases.
Explanation of the interlocks that secure the door and prevent opening.
Safety precautions related to water and electrical hazards during filling.
Description of the free steaming stage for air removal and warming.
Description of the heating stage to reach sterilization temperature.
Description of the main sterilization phase and its timing controls.
Details the heat-to-sterilise stage, including temperature and pressure controls.
Description of the sterilizing stage, including timing and temperature setpoints.
Guidance on selecting appropriate sterilization temperatures and times.
Detailed instructions on setting and adjusting cooling lock parameters for safe door opening.
Explanation of the pressure switch interlock, its function, and troubleshooting common issues.
Common problems associated with the water level sensor and their potential causes.
Explanation of overheat faults, their causes, and system behavior.
Detailed explanation of the optional load sensed timing feature and its benefits.
How cooling lock settings are influenced by the load sensed timing function.
Common issues encountered with the printer and their solutions.
General guidelines for maintaining the cleanliness of the steriliser.
Specific steps for cleaning the internal chamber of the steriliser.
Explanation of the safety interlocks that secure the door during operation.
Details the systems in place to protect against overheating.
Procedure for manually unlocking the door in emergency situations.
Steps for removing the steriliser handle.
Detailed steps for using the tool to manually unlock the door.
Recommended maintenance schedule including safety valve testing and gasket checks.
Instructions for cleaning the steriliser chamber and inspecting sensors.
Guidance on using temperature recorders for commissioning and program setup.