EasyManua.ls Logo

ATAGO DR-M2 User Manual

ATAGO DR-M2
4 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
P-SOP0803-1410-E01
1
/4
Standard Operating
Procedure SOP
Multi-Wavelength Abbe Refractometer
DR-M2
Preface
To ensure proper care, operation and many years of accurate measurement, the Standard
Operating Procedure (SOP) document for the MultiMulti-Wavelength Abbe Refractometer,
DR-M2, is provided herewith.
The following instructions should be used as an example. Based on the steps provided,
write up your companys own original SOP for the operation of the Multi-Wavelength Abbe
Refractometer, DR-M2.
To ensure proper operation, ensure that the main unit of the refractometer and the light
source unit are set up and connected. Also, the 589nm interference filter must be inserted
into the light source unit.
The standard operation procedure document of Multi-Wavelength Abbe Refractometer,
DR-M2
1. Operation Manual
1 Connect a circulating constant temperature bath to the refractometer, and power on.
2 Connect the AC power supply cable of the refractometer to the AC outlet.
Turn on the power supply switch of the DR-M2 and light source unit. The DR-M2
display should indicate [1] and the temperature. The Light source unit display
indicates the wave length and ON/OFF of the LAMP.
3 Drop about 0.1ml of distilled water onto the main prism surface, and gently close the
secondary prism. Set the light guide of the light source unit so that its relative position
is directed at the lighting window of the secondary prism. Press the LAMP key of the
light source unit. The LAMP indication on the display of the light source unit
changes toON. Press the SET key of DR-M2. Look at the scale through the eyepiece
and turn the measurement knob so that the boundary line is centered at the
intersection point of the cross hairs. Confirm that the temperature is 20.0(±0.2). The
display will indicate the refractive index of the sample being measured.
4 Press the SELECT key to display [1] in the mode indicator. Set the function mode
number to [4] with the key. Press the SET key and the refractive index at the
present boundary line blinks in the display. After confirming that the boundary line is
centered with the intersection point of the cross hairs, press the SET key again.
Press the SELECT key and reset the function mode number to [1] from [4] with the
key. Wipe off the distilled water with a soft, lint-free tissue paper.
5 When measuring a liquid sample, place approximately 0.1ml of the sample onto the
main prism surface, and gently close the secondary prism. Insert the proper
interference filter into the interference filter insertion slot of the light source unit. Set
the light guide of the light source unit so that its relative position is directed at the
lighting window of the secondary prism. Look at the scale through the eyepiece and
turn the measurement knob so that the boundary line is centered at the intersection
point of the cross hairs.
6 The value indicated as the boundary line crosses the intersection point of the cross
hairs is the measurement value of the sample. Wipe off the sample with a soft,
ling-free tissue paper.
7 When measurement is completed, be sure that any sample on the prism surface and
surrounding areas are completely wiped off with tissue paper moistened with water,
then wipe dry with tissue.
8 Turn off the power supply switch of the DR-M2. Confirm that nothing is indicated on
the display section.
9 If used, turn off the power supply switch of the temperature bath.

Other manuals for ATAGO DR-M2

Question and Answer IconNeed help?

Do you have a question about the ATAGO DR-M2 and is the answer not in the manual?

ATAGO DR-M2 Specifications

General IconGeneral
ModelDR-M2
Resolution (Brix)0.1%
Minimum Indication (RI)0.0001
Measurement Accuracy (RI)±0.0002
Light SourceLED
Wavelength589 nm
DisplayLCD
OutputUSB
CategoryRefractometer
Sample Volume0.3 mL
Power SupplyRechargeable battery
Temperature CompensationAutomatic

Summary

Preface

Operation Manual

Daily Check

Periodical Inspection

Inspection by Test Piece

Details how to inspect the DR-M2 using a test piece and distilled water for accuracy.

Record of Periodical Inspection

Table for recording results from periodical inspections of the refractometer.

Calibration

Maintenance

Troubleshooting and Malfunction Handling

Check Procedure

Steps to check for technical malfunctions by performing checks and measurements.

Proper Care for Malfunctions

Guidance on how to correct operational malfunctions based on their cause.

Establishment and Record of Revision

Related product manuals