Do you have a question about the ATAGO MASTER Series and is the answer not in the manual?
Identifies and illustrates the key components of the refractometer for user understanding and operation.
Check prism cleanliness, measure standard solutions, and verify measurement values against expected results.
Order Calibration Certificates based on ISO quality system; HACCP or GMP certification is optional.
This document describes the ATAGO MASTER-PX series refractometer, a device designed for measuring the refractive index of solutions, primarily expressed in Brix percentage. The series includes models such as the MASTER-53P α, MASTER-10P α, MASTER-20P α, and MASTER-P α, each with different measurement ranges.
The MASTER-PX series refractometers are optical instruments used to determine the concentration of soluble solids in a solution by measuring its refractive index. The Brix scale, based on sucrose and water solutions, indicates the total concentration of soluble solids. For samples like cutting oils or industrial fluids, a conversion chart may be needed to relate Brix percentage to total concentration. The device features Automatic Temperature Compensation (ATC), which automatically adjusts the internal scale to account for changes in ambient temperature, eliminating the need for manual temperature correction charts. This ensures accurate readings even when the ambient temperature fluctuates.
The MASTER-PX series offers various models with specific measurement ranges and accuracies:
All models feature Automatic Temperature Compensation (ATC). The protection class for the main unit (excluding the eyepiece) is IP65 Water Resistant, meaning it is dust-tight and protected against water jets.
Dimensions and Weight:
The device has received several design registrations and awards, including the Grand Prize of Invention Award (MEXT), JPO Commissioner Encouragement Prize, and Good Design awards in 2007 and 2009.
Quick and Easy Sampling: To use, apply approximately 0.3mL of sample to the front or rear end of the sample stage. Tilt the refractometer slightly to allow the sample to spread evenly over the prism. This method allows for quick readings, especially when measuring many samples, and is suitable for low-viscosity liquids.
Measurement Procedure:
Automatic Temperature Compensation (ATC): The ATC feature automatically adjusts the internal scale based on ambient temperature, eliminating the need for manual temperature correction. For accurate readings, the refractometer and sample should be at the same ambient temperature. When measuring heated or refrigerated samples, allow them to conform to ambient temperature for 1 to 2 minutes before placing on the prism.
Calibration: Periodic calibration is recommended. To calibrate, place purified water on the prism at an ambient temperature of 20°C. View the scale through the eyepiece and adjust the scale using the scale adjustment screw located on the underside of the device until the reading is correct.
Cleaning:
Eyepiece Maintenance: If moisture accumulates within the eyepiece, obstructing the view:
Storage: Store the unit in its original storage case in a dry, room temperature (0-40°C) environment away from direct sunlight.
Precautions:
Warranty and Repair: The MASTER-PX series is warranted for one year from the date of purchase against manufacturer defects in materials or workmanship. The prism and sample stage are excluded from the warranty. The warranty is void if the unit is disassembled by unauthorized personnel, misused, abused, used/stored in improper ambient conditions, immersed in liquid, or dropped. Service fees apply for repairs after the warranty period. The prism is considered a consumable item, and its replacement will incur a charge. All instruments sent for repair are subject to a possible inspection fee.
| Brand | ATAGO |
|---|---|
| Model | MASTER Series |
| Category | Measuring Instruments |
| Language | English |