Revision A
13
Options: Off
On
Default: Off
Enables or disables the audible portion of the 2H alarms. The unit parameter is factory set to off.
When L1 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting. Press the <Δ> (Up) and <∇> (Down) function keys to change the displayed setting
between ON and OFF.
Press <ENTER> to set the selection and return to the parameter select menu. [L1] will be
displayed.
L2 – 0% Sample Time
Options: 5 - 120
Default: 10
Establishes the period over which the 0% aerosol sample is average.
When [L2] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <Δ> (Up) and <∇> (Down) function keys to change the displayed setting between 5 and
120. Press <ENTER> when done. L2 will be displayed.
L3 – 100% Sample Time
Options: 5-120
Default: 10
Establishes the period over which the 100% aerosol sample is averaged.
When [L3] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <Δ> (Up) and <∇> (Down) function keys to change the displayed setting between 5 and
120. Press <ENTER> when done. [L3] will be displayed.