L4 – Decimal Places to be displayed
Options: 3 or 4
Default: 4
This parameter is used to select either 3 or 4 decimal places on the % Leakage Display.
When [L4] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <Δ> (Up) and <∇> (Down) function keys to move the displayed setting between 3 and 4
on the display (3=0.001%, 4=0.0001%).
Press <ENTER> to set the selection and return to the parameter select menu. [L4] will be
displayed.
L5 – Aerosol Reference Selection
Options: DOP, PAO, USER, HIGH
Default: DOP
When [L5] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Note: Refer to section 4.5, page 20 “changing the reference” to change the selection or
reestablish internal reference.
See A-7 for correction factors to be used in conjunction with other challenge aerosols.
Press <ENTER> to set the selection and return to the parameter select menu. [L5] will be
displayed.
L6 – Hour Meter
Options: N/A
Default: N/A
Displays the total running time for the unit.