Press <ENTER> to set the selection and return to the parameter select menu. [L9] will be
displayed.
L10 – % Leakage Display Intensity
Options: 1-8
Default: 6
[L10] is used to adjust the intensity of the green LED display in the % Leakage display.
When [L10] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <Δ> (Up) and <∇> (Down) function keys to move the displayed setting between 1 and 8
on the display.
Press <ENTER> to set the selection and return to the parameter select menu. [L10] will be
displayed.
L11 – 0% Purge Time
Options: 0 - 120
Default: 0
Set the length of time for the system to run before taking reading when calculating the zero value.
When [L12] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <Δ> (Up) and <∇> (Down) function keys to change the displayed setting between 0 and
120 seconds. Press <ENTER> when done. [L12] will be displayed.
L12 – 100% Purge Time
Options: 0 - 120
Default: 0
Set the length of time for the system to run before taking reading when calculating the 100%
value.
When [L12] is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.