16 AVACEN 100 User Guide 171208
21. USER SETTINGS
There are some settings that can be made on the AVACEN 100 that will be stored. Even if power
is removed, the settings are kept. The following User Settings can be made on the AVACEN 100:
• Language Selection
• Beep Tone Style
• Default Therapy Time
• Default Therapy Temperature
In addition, there are some settings that can be used to help diagnose problems with the device:
• Display Error Messages
• Measure Hand Pad Force
• Run Self-Sealing Test
All of these features are accessed in the User Settings menu. To access User Settings, begin by
turning the AVACEN 100 Off. Press and hold down the ▼ button while turning the device On
with the button. The display will show "User Settings".
Moving through the different User Settings is done using the front panel buttons:
◄ or ► Press to scroll back or forward through menu selections.
▲ or▼ Press to scroll up or down through menus and options.
✔ To store the selected setting.
◄ Go back to the previous (higher level) menu.
After you are done making changes to any of the settings, press ◄ one or more times until the
display shows "User Settings". Then press the button to return to normal operation.
While in the "User Settings" menu, press ►, then press ▲ or▼ one or more times to go to the
following menus:
A. Language Selection
Messages can be displayed in multiple languages on some
device versions. Press ► to enter this menu, then press ▲
or▼ to choose the desired language. Press ✔ to select the
desired language. If this menu is not available, then your
device only supports English.
B. Beep Tone Style
There are two options, "Original" and "Low Tones". Press
► to enter this menu, then press ▲ or▼ to choose a style
of Beep Tone. A sample of the tones will sound. Press ✔
to select the desired Beep Tones.
C. Default Therapy Time
The default setting for therapy time is 15 minutes. At any
time before the start of a therapy session, the therapy
time may be changed from 15 minutes to 10, 20, 25, or 30
minutes. If you find that you a usually selecting a time
other than 15 minutes for therapy, the default therapy
User Settings:
Change Display
Language
Beep Tone
Select
Default Therapy
Time/Temp
Set Default
Therapy Time