• Problem: Displayedweightuctuatesrandomly
• Solution: Make sure your workspace is stable
and free of any air currents, vibration, and static
electricity. Nearby radio wave emitting devices can
causeuctuations.
• Problem: Displayed weight is inaccurate
• Solution: Perform a calibration adjustment. If
you are still having trouble, contact our warranty
department.
IX. Error Codes
1. OUTZ - Zero point has drifted out of the acceptable
range. Try to perform a calibration adjustment. If the
problem persists, the load cell may be permanently
damaged due to overloading.
2. O-Ld - Overload. The maximum capacity has
been exceeded. Remove the extra weight to avoid
permanently damaging the load cell.
3. Lo - Batteries are low. Replace the batteries.
X. Specications
CUBE-100 CUBE-650
Max. Capacity 100g 650g
Readability 0.01g 0.1g
Min. Weight 0.03g 0.2g
Power 2 x 1.5V “AAA” Alkaline
Dimensions 3.1 x 3.1 x 0.7 in.
Platform Dimensions 3.1 x 3.1 in.
Warranty 10 Year Limited Warranty