SC-Series Manual
Model Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
SC-501,
SC-501A
0.02g
200g,
400g
OIML M2
SC-2KG,
SC-2KGA
0.2g
1000g,
2000g
OIML M2
VIII. Adjustment
Do Not Proceed Without Proper Weights (see above)
1. Turn the scale ON.
2. Press and hold the MODE key until the display
shows “CAL”, then release.
3. Make sure no weight is on the platform and press
MODE again.Thedisplaywillash“CAL”followed
bytherstrequiredcalibrationweight.
4. Placetherstrequiredweightontheplatform.
5. Oncetherstweightisconrmed,thedisplaywill
ashthesecondrequiredcalibrationweight.Place
the required weight on the platform.
6. Oncethesecondweightisconrmed,thedisplay
will show “PASS” and return to normal weighing
mode. Calibration is complete!
IX. Troubleshooting
• Problem: Scale will not turn on
• Solution: Check the batteries and battery contacts.
• Problem: Displayedweightuctuatesrandomly
• Solution: Make sure your workspace is stable and