Instruction for use
12
En
PRO-36
e. After measurement, the monitor will turn off automatically after 1 minute of no operation. Alternatively, you can press the “START”
button to turn off the monitor manually.
f. During measurement, you can press the “START” button to turn off the monitor manually. The device then immediately lowers the
cuff-pressure automatically, and you will hear at the time “B.Well wishes you all the best – stay healthy!”
Note:
Please consult a health care professional for interpretation of pressure measurements.
Note:
The monitor can memorize the last result. If you change the batteries, the last result may be lost.
7.8. Pulse Arrhythmia Detection
Appearance of the Arrhythmia indicator
The appearance of the symbol
signifi es that a certain pulse irregularity was detected during the measurement. The result can vary
from your normal blood pressure. As a rule this is not a cause for concern; however, if the symbol
appears more frequently
(e.g. several times per week on measurements performed daily) or if it suddenly appears more often than usual, we recommend you
inform your doctor. Please show your doctor the following explanation:
Information for the doctor on frequent appearance of the Arrhythmia indicator
The device is an oscillometric blood pressure measuring device that also analyses the pulse frequency during measurement. The
accuracy of this device has been clinically validated. If pulse irregularities occur during measurement, the arrhythmia symbol
is displayed after the measurement. If the symbol appears more frequently (e.g. several times per week on measurements
performed daily) or if it suddenly appears more often than usual, we recommend the patient to seek medical advice.
The device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.
7.9. Technical alarm description
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if the determined blood pressure (systolic or diastolic) is outside
the rated range specifi ed in part SPECIFICACIONS. In this case, you should consult a physician or check if your operation violated the
instructions.
The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or inactivated. This alarm condition
is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically after about 8 seconds in
the meantime, device reads out the possible causes..