4 3015-9005 Rev 1
WARNING: Except for replacement of
consumables such as sensors, filters, and
battery, this analyzer should only be
opened and/or serviced by authorized
Bacharach personnel. Failure to comply
may void the warranty.
WARNING: Do not store instrument or
its sensors with solvents or products that
contain solvents.
IMPORTANT: Equipment Damage
Hazards
A. Submerging the probe in liquid will
damage the pump.
B. Exposing the probe to pure refrigerant
will severely reduce the life or destroy
the sensor. Life of the sensor is
directly proportional to the amount of
refrigerant that it is exposed to.
C. Exposure to high concentrations of
refrigerant may require adjustment of
sensor heat. Refer to section 3.1.
1.4. Specifications
Spec Description
Battery
Type 12V lead-acid battery
Life 3 hours (typical)
Charging
3-4 hours with wall
adapter (included)
Sensor
Type Heated diode
Life
1 year (typical)
Calibra-
tion
As needed
(see Section 3.1)
Flexible Probe
4.5 ft. (1.4 m) length with
built-in filter
Response 1 second
Flow Rate
90 cc/min @ small setting
(typical)
Refrigerants
Detected
[2]
All CFCs, HFCs, and HCFCs