BANDELIN Sonopuls HD 4050 - User Manual

BANDELIN Sonopuls HD 4050
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Ultrasonic Homogeniser HD 4100
7
Ultrasonic homogenisers
51376g GB/2018-04
User Instructions
Valid for:
HD 4050 Volume: 0.5 ml - 100 ml
HD 4100 Volume: 2 ml - 200 ml
HD 4200 Volume: 5 ml - 1000 ml
HD 4400 Volume: 100 ml - 3000 ml

Table of Contents

Other manuals for BANDELIN Sonopuls HD 4050

Questions and Answers

  • C
    Connie SnyderAug 15, 2025
    What to do if my BANDELIN Sonopuls HD 4050 device cannot be switched on?
    • B
      Benjamin Adams Jr.Aug 15, 2025
      If your BANDELIN Laboratory Equipment device isn't switching on and the display is blank, start by ensuring the plug connection is firmly in place. Inspect the mains cable for any damage, and if needed, replace it. Also, check the device fuses located in the cold device socket on the rear side of the generator and replace them if they are tripped. Finally, replace the main fuse if it is faulty. These issues may arise from no power, a loose or faulty mains cable, a tripped device fuse, or a faulty main fuse.
  • A
    Amy FosterAug 21, 2025
    Why is my BANDELIN Laboratory Equipment heating up significantly near the mounting surfaces?
    • H
      hughessamanthaAug 21, 2025
      If you notice significant heating near the mounting surfaces between the ultrasonic converter and standard/booster horn, or between the standard/booster horn and probe of your BANDELIN Laboratory Equipment, it could be due to the sonicating parts not being mounted firmly enough or the mounting surfaces being soiled. To resolve this, dismount the respective parts, clean the surfaces, and firmly retighten them.
  • J
    Jennifer KelleyAug 25, 2025
    How to fix 'No USC' error on BANDELIN Sonopuls HD 4050 Laboratory Equipment?
    • C
      cristinaortizAug 26, 2025
      If you're getting a 'No USC' error on your BANDELIN Laboratory Equipment, it means the ultrasonic converter (UW) isn't connected to the ultrasound generator (GM). To fix this, connect the UW to the horn/probe on the GM. The GM should automatically recognize which UW has been plugged in.
  • M
    michelle27Aug 28, 2025
    What to do if frequency synchronisation malfunctions in BANDELIN Laboratory Equipment?
    • G
      Gabriel PenaAug 28, 2025
      If the frequency synchronisation is malfunctioning and the frequency display is not possible on your BANDELIN Laboratory Equipment, check whether the probe has been mounted correctly and is sufficiently tight. Inspect the horn/probe contact surfaces, horn/probe threads, and probe radiation surface for any damage. If needed, replace the UW/probe and then check its function using the 'probe check' or 'scan frequency' functions. After successfully completing the frequency scan, do not restart operation until initialisation is complete. In the event of failure, contact BANDELIN.
  • N
    nwrightAug 31, 2025
    What to do if frequency setting is not possible on BANDELIN Laboratory Equipment?
    • M
      Melissa NguyenAug 31, 2025
      If you're unable to set the frequency on your BANDELIN Laboratory Equipment, perform a frequency scan. If the error persists, contact BANDELIN.
  • K
    Kathy AustinSep 4, 2025
    What to do if there is no power output or amplitude setting is not possible on BANDELIN Laboratory Equipment?
    • V
      vickie50Sep 4, 2025
      Error correction for no power output or inability to set the amplitude on your BANDELIN Laboratory Equipment should only be carried out by BANDELIN. Send in the complete ultrasonic homogeniser.
  • J
    Jordan LawsonSep 7, 2025
    What to check if mains voltage has fallen short of on BANDELIN Sonopuls HD 4050 Laboratory Equipment?
    • R
      richardwalkerSep 7, 2025
      If the mains voltage has fallen short on your BANDELIN Laboratory Equipment, check the mains connection and the fuse.
  • D
    douglas08Sep 10, 2025
    What does 'unknown command' mean for BANDELIN Laboratory Equipment?
    • L
      Lance FrederickSep 12, 2025
      To address an unknown command or communication error with your BANDELIN Laboratory Equipment, observe the application information 'Protocol and command set for remote operation' and check commands.
  • B
    Brenda JenningsNov 17, 2025
    Why is there little or no ultrasonic power in my BANDELIN Sonopuls HD 4050?
    • C
      chavezmarkNov 17, 2025
      If you're experiencing little or no ultrasonic power with your BANDELIN Laboratory Equipment, several factors could be at play. First, verify that the connection from the ultrasonic converter to the standard/booster horn, or from the horn to the probe, is secure. If not, separate the parts, clean the surfaces, and firmly screw them back together using the provided tool. Also, check the horn, probe, and threaded pins for any cracks, and replace them if necessary, ensuring clean and flat surfaces. If liquid has penetrated between the standard/booster horn and probe, dismount the probe, clean the mounting surfaces and threads, dry and check for evenness, then remount and tighten the probe. If the threaded pin attachment on the titanium plate is defective, replace the titanium plate. Check the...

BANDELIN Sonopuls HD 4050 Specifications

General IconGeneral
TypeUltrasonic Homogenizer
Frequency20 kHz
Output Power50 W
Nominal Voltage230 V
PulseYes
AmplitudeAdjustable
Amplitude Range20 - 100%
Voltage230 V
Probe Diameter Options22 mm