Bante900P Portable Multiparameter Water Quality Meter
21
ORP Calibration Report
This program lets you check the millivolt offset in the ORP mode.
1. Press MR key in the ORP measurement mode, the display shows “LOC/P-01”.
2. Press or key until the meter shows “ELE/P-02” (Electrode Diagnosis).
3. Press ENTER key to confirm, the meter shows the last calibration date (Format: mm-dd).
4. Press key, the meter shows “OFS” (Offset Potential).
5. After the browsing, press MEAS key, the meter returns to measurement mode.