Bante900P Portable Multiparameter Water Quality Meter
27
1.4 Dip the conductivity electrode into the calibration solution, the meter automatically shows current calibration standard (e.g., 141S/cm).
1.5 Press ENTER key to confirm, “Calibration” indicator begins flashing.
1.6 Wait for the measured value to stabilize, the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed.
Enter