Radiologic:
Grade 1 or 2 distortion on chest X-ray.
Pinch-o should be evaluated for degree of severity prior to
explantation. Patients indicating any degree of catheter distortion
at the clavicle/rst rib area should be followed diligently. There are
grades of pinch-o that should be recognized with appropriate chest
x-ray as follows:
3,4
Grade Severity Recommended Action
Grade 0 No Distortion No Action
Grade 1 Distortion present without
luminal narrowing
Chest x-ray should be taken every one to three
months to monitor progression of pinch o
to grade 2 distortion. Shoulder positioning
during chest xrays should be noted as it can
contribute to changes in distortion grades.
Grade 2 Distortion present with luminal
narrowing
Removal of the catheter should be
considered.
Grade 3 Catheter transection or fracture Prompt removal of the catheter.
Cautions
Carefully read and follow all instructions prior to use.
Federal (U.S.A.) law restricts this device to sale by or on the order of a
physician.
Only qualied healthcare practitioners should insert, manipulate and
remove these devices.
Precautions
Follow Universal Precautions when inserting and maintaining the
catheter.
Follow all contraindications, warnings, cautions, precautions and
instructions for all infusates as specied by its manufacturer.
Use aseptic techniques whenever the catheter lumen is opened
or connected to other devices. Povidone-iodine is the suggested
antiseptic to use with this device and components. Acetone and
tincture of iodine should not be used because they could adversely