Cimarec
Stirring Hot Plates
OPERATION MANUAL
AND PARTS LIST
Model #
Top Plate Size Voltage
SP131014 4x4 100
SP131015 4x4 120
SP131010-33 4x4 220-240
SP131324 7x7 100
SP131325 7x7 120
SP131320-33 7x7 220-240
SP131634 10x10 100
SP131635 10x10 120
SP131630-33 10x10 220-240
LT1313X1 • 10/31/06

Questions and Answers

  • K
    kathrynbrownAug 9, 2025
    What to do if Barnstead International Laboratory Equipment displays message E11?
    • M
      Maria CarterAug 9, 2025
      If your Barnstead International Laboratory Equipment displays message E11, it could indicate several issues: * A locked rotor condition: Try to free the locked rotor. * Motor failure: You may need to replace the motor. * Motor circuit failure: Consider replacing the control board. * Motor sensor failure: Replacing the motor sensor might be necessary.
  • C
    Christopher WallaceAug 19, 2025
    What to do if Barnstead International Laboratory Equipment displays message E02?
    • E
      Erik SmithAug 19, 2025
      If your Barnstead International Laboratory Equipment displays message E02, it could be due to: * A thermocouple short circuit: Replacing the element assembly (attached to the thermocouple) might help. * Element failure: In this case, replace the element. * Element circuit failure: Try replacing the control board.
  • R
    Ronnie MillerAug 29, 2025
    How to fix error E12 on Barnstead International SP131320-33?
    • J
      Jennifer HarrisAug 29, 2025
      If your Barnstead International Laboratory Equipment displays message E12, replace the control board.

Summary

Safety Information

Warnings

Provides crucial safety advice regarding electrical shock and personal injury.

Specifications

Unpacking and Installation

Operation

Setting the Stirring Speed

Step-by-step guide to adjust and control the stirring function.

Setting the Temperature

Instructions for adjusting and controlling the heating temperature.

Troubleshooting

Error Codes

Explains error messages and their potential causes and solutions.