Cimarec
Stirring Hot Plates
OPERATION MANUAL
AND PARTS LIST
Model #
Top Plate Size Voltage
SP131014 4x4 100
SP131015 4x4 120
SP131010-33 4x4 220-240
SP131324 7x7 100
SP131325 7x7 120
SP131320-33 7x7 220-240
SP131634 10x10 100
SP131635 10x10 120
SP131630-33 10x10 220-240
LT1313X1 • 10/31/06

Questions and Answers

  • M
    Mrs. Casey BrooksAug 8, 2025
    What to do if Barnstead International Laboratory Equipment displays message E11?
    • D
      Dorothy RyanAug 8, 2025
      If your Barnstead International Laboratory Equipment displays message E11, it could indicate several issues: * A locked rotor condition: Try to free the locked rotor. * Motor failure: You may need to replace the motor. * Motor circuit failure: Consider replacing the control board. * Motor sensor failure: Replacing the motor sensor might be necessary.
  • G
    Gregory BlairAug 18, 2025
    What to do if Barnstead International SP131630-33 Laboratory Equipment displays message E02?
    • A
      Amy SullivanAug 18, 2025
      If your Barnstead International Laboratory Equipment displays message E02, it could be due to: * A thermocouple short circuit: Replacing the element assembly (attached to the thermocouple) might help. * Element failure: In this case, replace the element. * Element circuit failure: Try replacing the control board.
  • J
    jorgechapmanAug 29, 2025
    How to fix error E12 on Barnstead International SP131630-33 Laboratory Equipment?
    • M
      mark74Aug 29, 2025
      If your Barnstead International Laboratory Equipment displays message E12, replace the control board.

Summary

Safety Information

Warnings

Provides critical warnings for electrical shock and personal injury to ensure safe usage.

Specifications

Principles of Operation

Unpacking and Installation

Operation

Setting the Stirring Speed

Details how to adjust the stirring speed using the STIR knob and the feedback control mechanism.

Setting the Temperature

Explains how to set temperature setpoints, the role of the digital display, and the 'CAUTION - HOT TOP' light.

Troubleshooting

Error Codes

Lists displayed error messages, their causes, and recommended solutions for technical service.

Wiring Diagram

Replacement Parts

Two Year Bumper to Bumper Warranty

Barnstead International SP131630-33 Specifications

General IconGeneral
TypeWater Purification System
ModelSP131630-33
BrandBarnstead International
TOC Level< 5 ppb
Resistivity18.2 MΩ·cm