29
“rmP.r” segments (ramp to a setpoint at
degrees per second, minute or hour) and
“rmP.t” segments (ramp to setpoint in a spe-
cific amount of time). See “Setting the
Segment Type” for an explanation on how to
set a ramp segment.
1. Press the PAGE button until you
reach the program list (ProG LiSt).
2. Press the SCROLL button until dis-
play reads, “rmP.U.”
3. Press the UP or DOWN button to tog-
gle between seconds, minutes and
hours.
dwL.U: Setting Dwell Units
Dwell units are time units which are used in
“dwELL” segments (amount of time to remain
at a specific temperature ). See “Setting the
Segment Type” for an explanation on how to
set a dwell segment.
1. Press the PAGE button until you
reach the program list (ProG LiSt).
2. Press the SCROLL button until dis-
play reads, “dwL.U.”
3. Press the UP or DOWN button to tog-
gle between seconds, minutes and
hours.
CYC.n: Setting the Number of Cycles
Set the number of times a group of segments
or programs are to be repeated by following
the steps listed below.
1. Press the PAGE button until you
reach the program list (ProG LiSt).
8 S