Chapter 4: Maintenance 27
4. Remove the cap from the new sheath cubitainer and retain it
to use as a cap when the waste container is full.
5. Insert the sensor probe at a 45-degree angle into the sheath
cubitainer and tighten the cap assembly.
6. Press Restart.
More information Flushing the BD LSRFortessa system (page 30)
Replacing the waste cubitainer (page 27)
Replacing the waste cubitainer
Introduction This topic describes the procedures for preparing a 10% bleach
solution and for replacing the waste cubitainer. You must replace
the waste cubitainer whenever it gets close to being full.
You can use BD FACSClean™ at full strength as a replacement for
10% bleach.
Preparing 10%
bleach solution
Prepare a 10% bleach solution using a high quality, particulate-
free bleach. The bleach solution will be used to store the sensor
probe of the waste cap assembly while you change the waste
cubitainer.
To prepare a 10% bleach solution:
1. Pour 45 mL of distilled water into a 500-mL beaker.
2. Carefully add 5 mL of undiluted bleach to the same beaker.
Caution! Replace and tighten all caps when
reassembling the containers and cubitainers. Loose
caps can result in fluid leakage.
Caution! Prepare a fresh 10% bleach solution each time
you change the waste cubitainer