Why BeaconMedaes Z MED Medical Air System fails to start?
- CConnor FlemingAug 21, 2025
Your BeaconMedaes Medical Equipment may fail to start due to several reasons: * The main power might be disconnected, so turn on the main power. * There could be a power failure, so restore power. * The main fuse or a fuse in the control circuit might be blown, so replace the fuse. * The overload tripped on the starter, so reset it and check for system overload. * The high-temperature switch may be activated, so allow the unit to cool, reset it, and check for over-temperature conditions. * Low oil pressure, so check the oil level. * A loose or faulty connection, so check and tighten all wire connections.