DENTAL X-RAY
OPERATOR’S
INSTRUCTIONS
· Wall Mount Type ................. WK
· Floor Mount Type ................ FK1/FK2
· Mobile Type ........................ FM
· Room Mount Type .............. RK
· Ceiling Mount Type ............. CK
WARNING
This X-ray unit may be dangerous to patient and operator unless safe exposure
factors, operating instructions and maintenance schedules are observed.
R
0197

Other manuals for Belmont BELRAY II 097

Questions and Answers

  • C
    Cindy GordonAug 3, 2025
    What to do if AVR voltage cannot be adjusted to 94.5 Vac in "Adj" mode in Belmont BELRAY II 097?
    • C
      Cynthia SmithAug 3, 2025
      If you can't adjust the AVR voltage to 94.5 Vac in "Adj" mode for your Belmont dental equipment, first, check the cables between the control box and head to ensure they are correctly connected. If the cables are fine, the power generating PCB or timer PCB might be defective. Try replacing the power PCB first. If the AVR voltage still cannot be adjusted after replacing the power PCB, replace the timer PCB.
  • S
    Sarah HillAug 18, 2025
    Why doesn't "FIN" appear in tube current confirmation mode on my Belmont BELRAY II 097 Dental equipment?
    • M
      Melanie GarzaAug 18, 2025
      If "FIN" doesn't appear in the tube current confirmation mode on your Belmont dental equipment, it could be due to the AVR voltage being too high or too low. If the voltage is outside the range of 94.5±0.5Vac, adjust the voltage using VR1. If the voltage is within the correct range, manually set the final h.XX and e.XX values by pressing P·C·F and the power switch to enter manual setting mode.

Summary

[1] INTRODUCTION

USER NOTICES AND SAFETY

Essential safety and operational notices for users, including warnings.

MOBILE TYPE (FM)

ROOM MOUNT TYPE (RK)

CEILING MOUNT TYPE (CK)

SUB CONTROLLER OVERVIEW

[3] FUNCTION OF CONTROLS

MAIN POWER SWITCH OPERATION

Explains the function and operation of the main power switch for unit energization.

EXPOSURE TIME ADJUSTMENT

Details how to adjust exposure time using increment and continuous switches.

TOOTH SELECTION SWITCHES

Explains how tooth selection switches automatically set exposure times based on tooth type.

FILM SPEED SELECTION

Explains how to select and program film speeds for optimal exposure settings.

DIGITAL IMAGING MODE

PATIENT SIZE SELECTION

EXPOSURE TIME DISPLAY AND WARNINGS

EXPOSURE SWITCH OPERATION

[8] INFECTION CONTROL AND CLEANING

INFECTION PREVENTION MEASURES

Outlines measures for preventing infection during operation, including glove use and barriers.

EQUIPMENT CLEANING PROCEDURES

Provides procedures for cleaning the X-ray unit safely and effectively.

[9] ERROR CODES

Belmont BELRAY II 097 Specifications

General IconGeneral
BrandBelmont
ModelBELRAY II 097
CategoryDental equipment
LanguageEnglish