Instruction Manual for HPA4 with 1.0.X Firmware REV B Page 53
AC Power-Entry and Fuse
Module
Input Voltage Range
Note: The HPA4 is equipped with a universal
power supply. There is no voltage selection
switch. AC voltage range is 100 to 240 VAC,
50 to 60 Hz.
Power Cord
Note: The AC power input uses a standard
IEC type connector. One USA-compatible
power cord is included with the HPA4. IEC
style power cords in country-specific
configurations are available in your locality.
Caution: Always use a grounded power
cord. The HPA4 is equipped with a
standard IEC power entry module. Use
an IEC power cord that is equipped with
the appropriate connector for your
location. Cords are available from your
dealer.
Fuses
Caution: For continued fire hazard
protection always replace the fuses with
the correct size and type (T 0.5A 250V, 5
x 20 mm Littelfus Slo-Blo
®
HXP218.500 or equivalent). The fuse
drawer includes two fuses. Always
replace both fuses at the same time.