2.4 Control Test
The purpose of control solution is used to check the performance of the BeneCheck
PLUS System. The system and strips perform adequately if the control test result
falls within the indicated control range listed on the test strip vial.
The system should be checked
When you get your meter for the rst time or begin using a new box of test
strips.
When you suspect that the meter or the test strips are not working properly.
Any time you had repeated unexpected test results of the blood glucose, uric
acid or total cholesterol.
Any time you drop or bump the meter.
You can check system performance regularly or anytime by using control
solution when you suspect that test results are inaccurate.