Chapter 9 Specications
System Accuracy
:
± 20% at Glucose level ≥75 mg/dL (4.17mmol/L)
±
20%
at Uric Acid ≥ 5 mg/dL (0.30 mmol/L)
±
20%
at Total Cholesterol ≥ 150 mg/dL
Principle
:
Glucose & Total Cholesterol - Electrochemical
Biosensor ,Uric Acid - Electrochemical Sensor
Calibration
:
Plasma-equivalent
Test Sample
:
Fresh Capillary whole blood
Measuring Time
:
About 10 seconds (Glucose), 15 seconds (Uric Acid),
30 - 40 seconds (Total Cholesterol)
Measuring Range
:
Glucose 20 - 600 mg/dL (1.1 - 33.3 mmol/L)
Uric Acid 3 - 20 mg/dL (0.18 - 1.19 mmol/L)
Total Cholesterol 100 - 400 mg/dL (2.59 - 10.35 mmol/L)
Sample Volume
:
About 1 - 1.5 µL (Glucose & Uric Acid)
About 10 µL (Total Cholesterol)
Hematocrit Range
:
30 - 55% (for Glucose, Uric Acid)
35 - 50% (for Total Cholesterol)
Strip Storage Condition
:
10 - 30ºC (50 - 86ºF)