Additional Information for Healthcare Professionals :
When venous whole blood is used for the test, it can be collected into
heparin-containing test tubes winthin 30 minutes after drawing. The results may
be 7% lower than a capillary sample.
Interferences : Acetaminophen, uric acid, ascorbic acid, dopamine,
gentistic acid, ibuprofen, methyl DOPA, salicylate, tetracycline, tolazamide,
tolbutamide (when at physiological or therapeutical levels) do not signicantly
affect the results.
Patients undergoing oxygen therapy may yield falsely low results.
In situations of decreased peripheral blood ow, examples would include but are
not limited to severe dehydration, in shock, or in a hyperosmolar state (with or
without ketosis), hypertension, the test results may be falsely low.
BeneCheck PLUS Glucose Test Strips are plasma calibrated, testing with serum
or plasma sample will give falsely high results.
BeneCheck PLUS Glucose Test Strips have not been validated for testing with