56
Important information
for hearing care professionals about
Tinnitus SoundSupport
Device description
Tinnitus SoundSupport is a module function that can
be enabled in the hearing aids by the hearing care
professional.
Maximum wearing time
The wearing time of Tinnitus SoundSupport will
decrease as you increase the level above
80 dB(A) SPL. The fi tting software will automatically
display a warning when the hearing aid exceeds
80 dB(A) SPL. See “Max wearing time indicator”
next to the tinnitus fi tting graph in the fi tting
software.
The volume control is deactivated
By default, the volume control for the sound
generator is deactivated in the hearing aid. Risk of
noise exposure increases when the volume control is
activated.
57
If the volume control is activated
A warning may be displayed if you activate the
tinnitus volume control in the “Buttons & Indicators”
screen. This occurs if the relief sound can be listened
to at levels that may cause hearing damage. The
“Max wearing time” table in the fi tting software
displays the number of hours the patient can safely
use Tinnitus SoundSupport.
· Note the max wearing time for each program for
which Tinnitus SoundSupport is activated.
· Write those values in the table: “Tinnitus
SoundSupport: Limitation on use”, in the back of
this booklet.
· Instruct your patient accordingly.