Replace ProWax miniFit filter
1. Tool
Remove the tool from the
shell. The tool has two pins,
one empty for removal and
one with the new wax lter.
2. Remove
Push the empty pin into the
wax lter in the speaker and
pull it out.
New lter
3. Insert
Insert the new wax lter using
the other pin, remove the tool
and throw it out.
4. Final
The speaker should look as
shown when you have
replaced the wax lter.
✻
If you use a mold or LiteTip, a hearing care
professional must replace the wax lter in the
speaker.
Old lter
New lter
Empty pin