12
IEC 61000-4-4
Surge
IEC 61000-4-5
No application
No application
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
No application
No application
Power frequency
magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC61000-4-6
No application
No application
Radiated RF
IEC61000-4-3
10 V/m
80 MHz 2, 7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz 2, 7 GHz
80 % AM at 1 kHz
NOTE U
T
is the a.c. mains voltage before application of the test level.
Table 3
Guidance and manufacturers declaration - electromagnetic immunity
Radiated RF
IEC61000-4-3
(Test
specifications
for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communicatio
ns equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Modulation
Distance
(m)
IMMUNITY
TESTLEVE
L
(V/M)
385
380-
390
TETRA
400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430-
470
GMRS 460
FRS 460
FM
±5 kHz
deviation
1 kHz sine
2
0.3
28
710
704-
787
LTE Band
13, 17
Pulse
modulation
217 Hz
0.2
0.3
9
745
780
810
800-
960
GSM800/900,
TETRA800,
CDMA850,
LTE Band 5
Pulse
modulation
18 Hz
2
0.3
28
870
930
1720
1700-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,
Pulse
modulation
217 Hz
2
0.3
28
1845
1970

Related product manuals