EasyManua.ls Logo

Beta LaserMike LaserSpeed LS9000 - Laser Safety Precautions

Beta LaserMike LaserSpeed LS9000
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LaserSpeed 9000 Operator Guide
Safety Information
Part No. 93296 / Drawing No. 0921-01419 Page 8 of 16 Revision F (Feb 2013)
The following system specifications will help establish appropriate safety
measures.
Maximum Laser Power
0.050 watt
Laser Wavelength
0.785 micrometer
Minimum Laser Spot Size
(Elliptical)
5 x 3 millimeters
Beam Divergence
0.5 milliradians
Pulse Rate
Continuous wave
Maximum radiance (power
divided by spot-size area)
0.050 Watt / 0.1 cm
2
[0.5 W/cm
2
]
Laser Safety Precautions
The laser beam in the optical gauge is very powerful and can permanently
damage eyes not protected by laser safety glasses. To avoid exposing
yourself to hazardous radiation, you must take these precautions:
Never look into the laser beam. If you must look at the beam,
view it from an angle and in the direction in which the beam is
travelling.
The beams emitted from the Model LS9000 are invisible to the
unaided eye. Return the beam shutter to the closed position
when the system is not in use or during setup.
Ensure that all direct reflections are blocked.
Remove all rings, watches, or jewellery from your hands when
working on or near the gaugecan cause hazardous reflections.
Never install the instrument at eye level.
Operate the system only with people who have been instructed
in laser safety.
Post warning signs and lights that are active when laser is
operating.
If your country or state has no regulations governing the safe use of lasers,
Beta LaserMike recommends that you follow the guidelines specified by the
American National Standard for the safe use of lasers (ANSI Z136.11986).
For a copy of this document, write to:
Laser Institute of America
13501 Ingenuity Drive, Suite 128
Orlando, Florida 32826
1-800-345-2737