Loading...
Phenomenon Basic EMC
standard
Immunity test levels
Home Healthcare
Environment
Radiated RF EM
field
IEC 61000-
4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-
4-3
Refer to table 3
Rated power
frequency
magnetic fields
IEC 61000-
4-8
30A/m
50Hz or 60Hz
Table 3 – Proximity fields from RF wireless
communications equipment
Test fre-
quency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare faci
-
lity environment
385 380-390 Pulse modulation 18Hz,
27V/m
Test fre-
quency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare faci
-
lity environment
450 430-470 FM, ±5kHz deviation, 1kHz
sine, 28V/m
710 704-787 Pulse modulation 217Hz,
9V/m
745
780
810 900-960 Pulse modulation 217Hz,
9V/m
870
930
1720 1700-1990 Pulse modulation 217Hz,
28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz,
28V/m
24