7.2 Disinfection
Device
Please comply with the generally applicable guidelines on disinfection when using the device on dif-
ferent persons. Do not submerge the device in disinfection solutions or any other fluids and make
sure that no fluids can enter the device.
Note
The blood glucose monitor is made of precision components. The accuracy of the measurements
and service life of the device depend on its careful handling:
Protect the device from impacts and do not drop it.
Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or
water, strong temperature fluctuations, direct sunlight and extreme cold.
Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile tel-
ephones.
8 what if there are problems?
Display messages on batteries and blood glucose measurement
No. Cause Solution
LP Batteries empty. Replace all batteries.
Ht
Temperature of the measuring environ-
ment, blood glucose monitor or test strip
above the permitted range.
Repeat the test using a new test strip as soon
as the measuring environment, blood glucose
monitor and test strips have reached room tem-
perature (+20 °C to +26 °C).
Lt
Temperature of the measuring environ-
ment, blood glucose monitor or test strip
below the permitted range.
Repeat the test using a new test strip as soon
as the measuring environment, blood glucose
monitor and test strips have reached room tem-
perature (+20 °C to +26 °C).
Err
Used or contaminated test strip inserted.
Insert unused test strip that has not expired.
Repeat the blood glucose measurement.
Err001
System error.
Remove batteries, reinsert batteries. Contact
our customer service if the problem persists.
Err002
Insucient amount of blood on the test
strip.
Repeat the measurement using a new test strip.
Err005
System error.
Remove batteries, reinsert batteries. Contact
our customer service if the problem persists.
Unknown error messages.
Remove batteries, reinsert batteries. Contact
our customer service if the problem persists.