CLASSIFICATION
1. Class of protection against electrical shock.
CLASS II EQUIPMENT
2. The degree of protection against electric shock.
APPLIED PART—TYPE B
3. Mode.
CONTINUOUS OPERATION WITH INTERMITTENT
LOADING
4. According degree of protection against ingress of water..
IPXO
FUSE REPLACEMENT
Occasionally power surges, etc. or normal age can result in a blown outer safety fuse located in the rear of
pump, adjacent to the power cord.
The safety fuse may be replaced by the user or caregiver, provided it is replaced with an identical type (Buss
MDL3 250V).
Prior to removal of fuse, disconnect power cord from socket. While pushing inward on fuse cap, turn
counter clockwise to release cap and remove fuse. After placing new fuse in cap slot, push cap and fuse in-
ward and turn clockwise to lock in place.
NOTE: The outer safety fuse is the only item serviceable by some one other than a Bio Compression Sys-
tems technician at the factory. Bio Compression Systems technicians have been trained specifically for the
manufacturing and repair of all Bio Compression Systems products.
NOTE: Having little or no “electromagnetic” or “radio frequency” signal sensitivity type components, these
devices neither generate nor are they affected by any of this type of interference. Further, their accuracy
remains consistent in the presence of such devices emitting this type of interference.
PUMP ENCLOSURE
The pump enclosures are constructed of “Cycolac” which is a trademark of General Electric.
UL FLAME RATING: Under file #E47016, The UL Test method of UL 94, @ 23°C, resulted in a Flamma-
bility Rating of (2.3 VO)
UNPACKING EQUIPMENT
SEQUENTIAL CIRCULATOR
By laying carton on its side, slide pump out with protective foam end caps still attached. After removing
pump from carton, foam end caps may be detached by gently pulling off each side.
NOTE: Be sure to SAVE carton and end caps for future transporting or shipping. When transporting, the
carton is equipped with “fold-up” handles for easy carrying.
SLEEVE/GARMENT
Remove sleeve from package and unroll tubing section which is permanently attached to garment with tub-
ing plug attached to the end.