Qualification Procedures | 115
To run the Absorbance Plate Test in KC4™:
1 Open the System menu and select Diagnostics|Run
Universal Plate Test
.
2 Select the Universal Plate and then click Run Test.
To run the Absorbance Plate Test in KCjunior™:
1 Open the Utilities menu and select Diagnostics|Universal
Plate Test
.
2 Select the Universal Plate ID and then click Run Test.
Test Failures
If any of the test parameters report as "FAIL," confirm that the standard
values on the test plate data sheet match the values on the printout. If not,
correct and retest. If the test still fails, contact Bio-Tek's Technical
Assistance Center (refer to
Chapter 1 for contact information). Please have
a copy of the test and the reader’s serial number available when you call.