This Manual Update contains changes to the ELx800™ Operator’s Manual.
These changes will be incorporated in the next full revision (Rev O) of the Manual.
Page 43, System Startup
Addition:
L
Important! Some ELx800 readers have custom programmed
software installed. Not all features of the software discussed
in this Operator’s Manual are available on custom instruments.
Please contact Bio-Tek’s Technical Assistance Center at 800-242-4685
if you have any questions about the software on your reader.
Page 119, Liquid Test 1
Recommendation:
After pipetting the diluted test solution into the microplate (step 3) and
before reading the plate, we strongly recommend shaking the plate for
four minutes. This will allow any air bubbles in the solution to settle and
the meniscus to stabilize. If a plate shaker is not available, wait 20 minutes
after pipetting the diluted test solution before reading the plate.
7331000 Rev N2
July 2005