SW41 SHORT SUCR 5-50% wv 11 Steps
Time/Angle/Speed
0:05/87/30
0:15/87/0
0:05/87/30
0:15/87/0
0:05/87/30
0:15/87/0
0:05/87/30
0:15/87/0
0:05/87/30
0:15/87/0
0:19/80/20
Note that the angle stays the same for all but the last step and that the speed alternates between 30
rpm and 0 rpm for 5 repeats: thus the name START-STOP. To simplify data entry, you will only need
to describe the three steps that are different. Each is stored in a separate memory (M1, M2, M3, etc.)
and then run in the desired order.
For example, M1 = 0:05/87/30, M2 = 0:15/87/0, M3 = 0:19/80/20. After programming each memory,
you will set the order of their appearance in your plan. For the 11 step run illustrated, the order is:
1 - 2 - 1 - 2 - 1 - 2 - 1 - 2 - 1 - 2 - 3
IMPORTANT: the time should be entered as 0:00 for any step that you want to leave indeterminate,
typically the last step which is stopped manually as the blue dye starts to penetrate the last bit of the
light solution. Here, since we are unsure what the mixing time for the 11th “polishing” step will be,
M3 should have time set to 0:00 (M3 = 0:00/80/20).
9. Entering the Memories:
You have finished describing the run, now enter the different memories:
Short Sucr 02-28% wv
EDIT NEXT DONE EXIT
Pressing the DONE key brings up the first Memory
Mem 1 0:05m 87.0° 30rpm
TIME NEXT MEM# DONE EXIT
To change the time, press TIME:
Time: 000 min. 05 sec.
DONE MIN. SEC.
05 is blinking. Press 0-7 to change the time to 7 sec, or MIN 0-0-6 to change minutes to 6 min. then
press DONE to return to the next entry, ANGLE:
Mem 1 0:07m 87.0° 30rpm
ANGL NEXT MEM# DONE EXIT
15