The display should show:
GRADIENT MASTER SD v1.5
CODE GRAD ROTR
To load new gradient run parameters, press the GRAD key. The display reads:
Downloading SD HEX file.
Rec loaded = 0 -> 2500
After a brief pause, the display starts counting up from 0 again:
Downloading SD CODE file.
Rec loaded = 0 -> 0400
The display then reads:
Downloading SD CODE file.
Wait,COPING Flash Page 1
After ~40 sec, the display reads:
FLASH Programmed, Reset
Wait,COPING Flash Page 1
6c. Now you’ll need to reload the rotor keys, since loading a new Grad List erases all keys you
had previously loaded. Repeat the startup routine and press ROTR:
GRADIENT MASTER SD v1.5
NEXT Rotor: SW28 LOAD
Use the NEXT key to scroll to the rotor you want to load and press LOAD.
Downloading SD HEX file.
Rec loaded = 0000-1300
The display informs you that the key has successfully been loaded:
Version 1.1
Enabled SW28 ROTOR
Repeat this procedure until all the keys are loaded. When you turn the unit off and on and go to the
GRADIENT MENU and press the LIST key, you will see all the rotors you have loaded.
6.!Turn the GM off and back on and you are ready to go.