6D. Step 4H
“Abandoned” vial left
in unit more than 15
seconds after thawing
process is completed
Error Code
7
Front: Flash full circle with
red light background once
and then display red light
background with error code 7
(seven LEDs at the top of the
circle: 13,14,15,1,2,3,4)
Four tone
sequence for
5 seconds,
then two tone
sequence every
15 seconds until
vial is removed.
Gently remove vial. Unit will emit a
single tone. Unit will display a single
light at the top, indicating that the unit
is ready for the next thaw.
6E. Step 4E-H
Vial fails to rise after
thaw process is
nished
Error Code
9
Front: Flash full circle with
red light background once
and then display red light
background with error code
9 (nine LEDs at the top of the
circle: 12,13,14,15,1,2,3,4,5)
(If in step 8, error code 7
displayed after 15 seconds)
Four tone
sequence for
5 seconds,
then two tone
sequence every
15 seconds until
vial is removed.
Grasp the vial cap only with ngers or
forceps and rock vial gently to dislodge
and remove from opening, taking care
not to open the cap. Unit will display
a single light at the top, indicating
that the unit is ready for the next thaw.
Ensure that the thermal pads lining the
internal heating blocks are intact and
undamaged. Note: If thermal pads are
not intact and/or damaged, contact
BioLife customer service. Avoid sticking
forceps or other foreign objects into the
opening.