Patient monitor Users manual
12
Delay of apnea alarm 10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
Alarm Indication
Blinking display of the data and parameters, text prompts,
Three levels of alarming: sound-light alarming, alarming
with blinked data and parameters, and that with text
prompts.
Recovery time after
defibrillation
<5s
A.6.3 NIBP
Standard IEC80601-2-30, ISO 81060-1, ISO 81060-2, SP10
Measurement way Automatic oscillometry
Measurement kinds Sys, Dia, Map,PR
Measurement range
mmHg
Adult
Sys
30~270 mmHg
Dia
10~220 mmHg
Map
20~235 mmHg
Pediatric
Sys
30~235 mmHg
Dia
10~220 mmHg
Map
20~225 mmHg
Neonatal
Sys
30~135 mmHg
Dia
10~110 mmHg
Map
20~125 mmHg
Cuff pressure range
0 mmHg to 280 mmHg
Resolution 1 mmHg
Pressure accuracy
Static:
Clinic:
±3 mmHg
Average error: ±5 mmHg, standard deviation: ≤8 mmHg
Unit
mmHg, kPa
Cuff auto deflation
The cuff will deflate automatically when power is off
or time of measurement is beyond 120 seconds(90
seconds for neonate) or the cuff pressure is beyond the
overpressure protection set by software and hardware.