Patient monitor Users manual
19
Accuracy ( 760mmHg
temperature is 25 )
0 mmHg to 40 mmHg, ±2 mmHg
41 mmHg to 70 mmHg, ±5% of reading
71 mmHg to 100 mmHg, ±8% of reading
101 mmHg to 150 mmHg, ±10% of reading
(when RR 80 rpm, all the range is ±12% of reading) Gas
temperature at 25
Total time of
system
response
3s
Sampling frequency
and accuracy of gas
Sampling frequency: 50mL/min
Accuracy: -7.5mL/min+15mL/min
Recovery time after
defibrillation
<5s
A.6.8 AG
Mainstream AG module (IRMA)
Measurement way Infrared spectrum
Measurement mode Mainstream,Sidestream
Measurement
parameters
CO
2
, N
2
O, agentISO, ENF, SEV, HAL, DES
Resolution 1%
Calibration
Zeroing recommended when changing Airway adapter. No
span calibration required for the IR bench.
Room air calibration of oxygen sensor performed
automatically when changing airway adapter (<5s)
Warm up time
Concentrations are reported and the automatic agent
identification is running within 10s. Full accuracy within 20s
for IRMA AX+.
Rise time
(When the flowing
speed is 10 L/min)
CO
2
≤ 90 ms
N
2
O ≤ 300 ms
ISO, ENF, SEV,HAL,DES ≤ 300 ms
Total system
Response Time
1s
Primary agent
threshold
0.15%. When an agent is identified, concentrations will be
reported even below 0.15% as long as apnea is not detected.