Patient monitor Users manual
36
Technical alarm 1displayed in the message area:
Alarm messages Cause Level
NIBP communication
error
NIBP measurement failure or
communication failure
Medium
Technical alarm 2displayed in the indicating area under NIBP MAP:
Alarm messages Cause Level
SELF-TEST FAILED Transducer or other hardware failure.
Low
LOOSE CUFF 1. Cuff is completely unwrapped.
2. The cuff is not connected.
3. Adult cuff used in neonate mode.
AIR LEAK Air leak in pneumatics, hose, or cuff.
AIR PRESSURE
ERROR
Unable to maintain stable cuff pressure,
e.g. kinked hose.
WEAK SIGNAL Very weak patient signal due to a loosely
wrapped cuff. The pulse of patient is too
weak.
RANGE EXCEEDED Measurement range exceeds module
specification.
EXCESSIVE MOTION
1.Too many retries due to interference of
motion artifact.
2.
Signal is too noisy during measurement,
e.g. patient has severe tremor.
3.Irregular pulse rate, e.g. arrhythmia.
OVERPRESSURE
SENSED
Cuff pressure exceeds the specified high
safety limit. Could be due to rapid
squeezing or bumping of cuff.
Medium
SIGNAL SATURATED
Large motion artifact that saturates the
BP amplifiers amplitude handing
capability.
Low
PNEUMATIC LEAK
Module reports Air Leakage failure while
in the Pneumatic Test mode.
SYSTEM FAILURE Module occurs abnormal processor event.
TIME OUT Measurement took more than 120
seconds in adult, 90 seconds in neonate
mode.
CUFF TYPE ERR Neonate cuff used in adult mode.
Prompt messagesdisplayed in the indicating area under NIBP MAP:
Messages Cause Level
NIBP Resetting NIBP measurement module is resetting
No alarm
Overpress Testing NIBP is testing Over-Pressure
Manometer Testing NIBP is testing Manometer